|
One elite premium rasp knife used for treatment but was not returned for evaluation.This is a reusable instrument.Due to unavailability, the allegation could not be fully confirmed.Definitive conclusions, investigation and evaluation were limited without physical evaluation.If objective evidence becomes available to assist with evaluation, the complaint will be revisited.Factors that may affect device performance include: device storage, device ability, surgical ability, procedure location and tissue condition.Instructions for use confirms instructions, precautionary statements and recommendations for proper use of product.Influences that could compromise product performance or integrity include: 1) current recommended maintenance.2) entanglement with other instruments or devices during use or cleaning.1.Compliant use of instrument.2.Careful cleaning and sterilization of product.3.Entanglement with other instruments or devices during use or cleaning.4.Pressure or excess torque applied.Per instructions for use: ¿these instruments are designed for repeated use with proper care and handling.¿ instrument tips may inadvertently be twisted or caught on other instruments or the basket during cleaning.This can lead to compromised strength.¿as with any surgical instrument, careful attention should be made to assure that excessive force is not placed on the instrument.Excessive forces applied to the instrument can result in the instrument¿s failure.Do not use these instruments as levers for manipulating hard tissue or bone.Excessive force should not be applied to the instrument when manipulating soft tissue, bone, or hard objects.Misuse of these instruments may result in bent distal tips or jaws; and dull or uneven cutting edges.¿ product met specifications upon release to distribution.
|
|
One 72201661 elite premium knife rasp instrument used for treatment, was returned for evaluation.This is a six year old reusable product.The complaint stated: ¿the device broke into three pieces.¿ this was confirmed.The instrument was fractured at both ends of the shaft.The shaft itself was visibly bowed.Per ifu: ¿these instruments are designed for repeated use with proper care and handling.¿ instrument tips may inadvertently be twisted or caught on other instruments or the basket during cleaning.This can lead to compromised strength.As with any surgical instrument, careful attention should be made to assure that excessive force is not placed on the instrument.Excessive forces applied to the instrument can result in the instrument¿s failure.Do not use these instruments as levers for manipulating hard tissue or bone.Excessive force should not be applied to the instrument when manipulating soft tissue, bone, or hard objects.Misuse of these instruments may result in bent distal tips or jaws; and dull or uneven cutting edges.¿ there are no other complaints for this lot.This allegation is considered an isolated incident.Product met specifications upon release to distribution.No root cause related to the manufacturing of this device was confirmed.
|