• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC BIVONA; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC BIVONA; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number 60P045
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/20/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation results: two bivona uncuffed neonatal tracheostomy tube were received without their original packaging inside a (b)(6) bag.The samples were returned in a used condition and were accompanied by their certificates of decontamination.Visual inspection was performed on the samples at 12" under normal conditions of illumination.It was observed that the samples had splits in their shafts and had exposed wires inside of them.In addition, the inspection revealed some contaminations.The device history record was reviewed and showed that this device met all manufacturing specification for product released for distribution.No issues were identified that would have impacted this event.The most probable root cause was that the damage occurred after the product left the manufacturing facility.
 
Event or Problem Description
It was reported that a smiths medical bivona uncuffed neonatal tracheostomy tracheostomy tube split through where the hub of the connector and shaft meet.The issue was discovered when the patient's mother was cleaning the tube she had just removed from the patient's airway.No adverse effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIVONA
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC
6000 nathan lane north
minneapolis MN 55442
MDR Report Key8764157
Report Number3012307300-2019-02136
Device Sequence Number16718360
Product Code BTO
UDI-Device Identifier15021312005769
UDI-Public15021312005769
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K912469
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number60P045
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/21/2019
Initial Report FDA Received Date07/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-