This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on july 5, 2019.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date).G4 (date received by manufacturer).G7 (indication that this is a follow-up report).H2 (follow-up due to additional information).H3 (device evaluated by manufacturer).H4 (device manufacture date).H6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).Method code #1: 10 - actual device evaluated.Method code #2: 11 - testing of device from same lot/batch retained by manufacturer.Method code #3: 3331- analysis of production records.Results code: 3259 - improper physical structure.Conclusions code: 4307 - cause traced to component failure.The affected sample was returned for evaluation, was checked to confirm that the arterial temperature port was non-functional.A retention sample from the same production/lot number combination was tested and found to be functioning properly.An engineering investigation and capa have been initiated to determine a definitive root cause.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
|