|
The reason for this revision surgery was due to post-operative injury of soft tissue and ligaments.The previous surgery and the revision detailed in this investigation occurred over 1 year and 1 month apart.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported components used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports associated with the product that may have contributed to the event.The device was within its respective expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to post-operative injury of soft tissue and ligaments.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's event.It seems that the event may have possibly occurred due excessive weight, excessive loads or torques, degenerative bone disease, poor tendon quality, patient non-compliance with medical instruction, patient activities or trauma.No additional information was submitted with the complaint regarding pre-existing conditions of the patient or any activities that may have contributed to the event and hence a definitive root cause cannot be determined.There are no indications of a product or process issue affecting implant safety or effectiveness.
|