|
Clinical assessment was completed based on the received medical records.The reported polymer leak, and intraoperative type 1a endoleak treated with a bare metal non endologix stent was confirmed from the still images.The patient experiencing hypotension and was successfully treated with an anaphylactic reaction protocol event is unconfirmed due to the lack of an operative notes.The contributing factors for the intra-operative type 1a endoleak was most likely the polymer leak, resulting in loss of seal around the sealing rings.The contributing factors for the polymer leak could not be identified.However, definitive device, user, or anatomy relatedness of this event could not be determined.The procedure related harms reported was intraoperative hypotension, that was successfully treated.The final patient status was reported being stable and discharged.As identified in fs-0012, the root cause for most polymer leaks is a material weakness adjacent to the polymer fill channel which may become compromised during pressurization with liquid polymer.Since this device remains implanted, endologix is unable to conduct product evaluation to conclusively ascertain root cause.However, based on the investigation associated with this fsn, this is the most likely cause associated with this event.On (b)(6) 2018, endologix issued a field safety notice (fs-0009) regarding the risk of polymer leak events with the ovation ix aortic body stent graft.On 06may2020, endologix issued a follow-up safety update (fs-0012) reaffirming treatment recommendations for patients who experience a polymer leak during implantation and providing updated information on the current rate of polymer leaks, the rate of clinical harms and root cause.The manufacturing lot evaluation confirmed all devices met specifications prior to release.The device was not returned as the device remains implanted therefore no evaluation was completed.No additional investigation of this reported event is planned however if additional information pertinent to the incident is obtained, a follow-up report will be submitted.Endologix will continue to monitor this complaint and similar complaints in the event further investigation is needed.Corrections: g1,2: contact office- contact has been updated h6: result code: remove code 3233.
|