Per the user facility's biomedical technician, this is for their back-up system.The field service representative (fsr) verified the broken latch.He replaced the latch.The unit operated to the manufacturer's specifications.The suspect part was returned to the manufacturer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
It was reported that during set-up of the device for a cardiopulmonary bypass (cpb) procedure, the centrifugal drive motor had a broken latch.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no delay, no blood loss, nor adverse consequences to the patient.
|