| Catalog Number |
0064330 |
| Medical Device Problem Code |
Output Problem (3005)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/26/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On june 26, 2019 immucor service engineers determined that immulink was mapped incorrectly causing the discrepant interpretation; the mapping problem was corrected at that time.Mapping is part of the customizable configuration that occurs during installation; the mapping is validated by the customer as being correct as part of the initial software installation.After installation, the customer does not have permissions to alter the mapping.Customer was notified of the findings and that the issue had been corrected.(b)(4).
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2019 a customer complained that immulink had reported an incorrect interpretation for a donor confirm (forward abo and rh); an ab positive sample was correctly interpreted on the neo analyzer but the result in immulink was displayed a positive.
|
| |
|