Model Number N560
Medical Device Problem Code Device Alarm System (1012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the unit had alarm defect.There was no allegation of patient death or serious injury.
 
Additional Manufacturer Narrative
Evaluation summary: one device was received and disinfected.Visual examination found damages to the external lower casing and the data label.Functional testing of the device found the battery was depleted, expired past its life cycle.During testing the device low battery alarms functioned as intended to indicate that the battery is low and does not charge properly.During maintenance the battery was replaced, and the unit passed all functional and electrical safety tests.The customer refused repairs and did not want the unit returned.Information has been added to the database and trends will continue to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NELLCOR
Common Device Name
OXIMETER
Manufacturer (Section D)
MEDIANA CO. LTD.
1650 1 2 donghwa ri munmak eu
kangwon do wonju si 00000
KR  00000
MDR Report Key8769181
Report Number2936999-2019-00457
Device Sequence Number19327221
Product Code DQA
UDI-Device Identifier10884521582538
UDI-Public10884521582538
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K021090
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN560
Device Catalogue NumberN560
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2019
Initial Date Received by Manufacturer 06/19/2019
Supplement Date Received by Manufacturer08/19/2019
Initial Report FDA Received Date07/08/2019
Supplement Report FDA Received Date09/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-