• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 5.5 HEALIX TI ANCHR W/OC +NDLS; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 5.5 HEALIX TI ANCHR W/OC +NDLS; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE Back to Search Results
Catalog Number 222242
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 06/06/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Initial reporter phone number: (b)(6).Udi: (b)(4).The expiration date is unknown.
 
Event or Problem Description
It was reported by the affiliate via cst that during a rotulian tendon reconstruction the 5.5 healix ti anchor with orthocord and needles was used with a punch starter and passed without problems.The surgeon began to repair the tendon with one of the threads, and when he went to take a top thread, the thread was sheared and frayed, which is broken, and the four strands of the anchor were released.Obviously, it was necessary to replace it for another one and as they did not have the same thread, a fastin 5.0 was used, and with this second healix to make the repair.The surgery was completed successfully.It is unknown if fragments were created.There was no surgical delay.No patient consequence.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.The complaint device was received and evaluated.Upon visual evaluation, it was found that the device was incomplete as it was missing the cover handle, suture card and the anchor.Only the healix driver and the handle were returned.The sutures were not returned; therefore, the complaint can't be confirmed.A root cause cannot be determined as the components did not return with the device for a physical evaluation.Furthermore, a non-conformance search was performed and no non-conformances were identified for this part number, lot number combination per qlik query executed on 6/10/2019.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.No nonconformances were identified for this part number, lot number combination per qlik query executed on 6/10/2019.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional information: (b5): event description.
 
Event or Problem Description
Additional information received from the affiliate reported the procedure was not an arthroscopic surgery.The affiliate also reported additional information is not available.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10: additional narrative: d10 h3, h6: the device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 HEALIX TI ANCHR W/OC +NDLS
Common Device Name
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key8769403
Report Number1221934-2019-57557
Device Sequence Number2540161
Product Code HWC
UDI-Device Identifier10886705007882
UDI-Public10886705007882
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K082282
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number222242
Device Lot NumberL303271
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/03/2019
Initial Date Received by Manufacturer 06/08/2019
Supplement Date Received by Manufacturer07/19/2019
08/30/2019
09/03/2019
Initial Report FDA Received Date07/08/2019
Supplement Report FDA Received Date08/16/2019
08/30/2019
09/04/2019
Patient Sequence Number1
-
-