BIOMET MICROFIXATION 90° CONTRA ANGLE DRIVER; SCREWDRIVER Back to Search Results
Model Number N/A
Medical Device Problem Codes Defective Device (2588); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Information (3190)
Date of Event 03/12/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimmer biomet complaint (b)(4).Medical product: biomet microfixation temporary fixation screw catalog #: 76-0017, lot #: 703610.Complaint sample was evaluated and the reported event was confirmed.Functional testing was done on the driver by rotating the knob, which showed that the driver was sticking during rotation.The driver was disassembled for further inspection and it was found that the internal gears were stripped and metal shavings fell out during disassembly.Dhr was reviewed and no discrepancies were found.Investigation results concluded that the reported event was due to the gears being stripped, likely resulting from torque in excess of what is required to fixate the screw.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2019-00356 & 0001032347-2019-00217-1.
 
Event or Problem Description
It was reported by the sales associate that a temporary fixation screw became stuck in one driver and another driver was hard to turn.No adverse events have been reported as a result of the malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
90° CONTRA ANGLE DRIVER
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
MDR Report Key8770061
Report Number0001032347-2019-00355
Device Sequence Number14119686
Product Code HRS
UDI-Device Identifier00841036123130
UDI-Public00841036123130
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K121589
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number24-1189
Device Lot Number548890
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2019
Initial Report FDA Received Date07/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-