DEPUY IRELAND - 9616671 RECLAIM DISTAL TAPERED 14X140; RECLAIM IMPLANT : HIP FEMORAL STEM Back to Search Results
Model Number 1976-14-140
Medical Device Problem Codes Device-Device Incompatibility (2919); Difficult to Open or Remove Packaging Material (2922)
Health Effect - Clinical Codes Not Applicable (3189); No Code Available (3191)
Date of Event 06/17/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint : (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Product details: reclaim stem 14mm x 140mm straight stem.1976-14-140 lot h64477 date: 17/6/2019.Issue: femoral canal prepared as per surgical technique.Above femoral stem implanted as per surgical technique.When surgeon went to remove taper protector sleeve it would not disengage.Proximal aspect of sleeve eventually broke leaving distal part in situ.Stem then had to be removed and another implant opened and implanted in its place.Case length was extended by approx 15 minutes.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned.The report has been confirmed using customer provided photographs.The root cause is attributed to design.Corrective action was established via the depuy corrective and preventive action system.Additional corrective action is not indicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H6 patient code: no code available (3191) used to capture the insufficient information and surgery prolonged.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECLAIM DISTAL TAPERED 14X140
Common Device Name
RECLAIM IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key8771704
Report Number1818910-2019-97829
Device Sequence Number2038898
Product Code KWA
UDI-Device Identifier10603295076995
UDI-Public10603295076995
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K102080
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1976-14-140
Device Catalogue Number197614140
Device Lot NumberH64477
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/18/2019
Supplement Date Received by Manufacturer08/21/2019
09/22/2020
Initial Report FDA Received Date07/09/2019
Supplement Report FDA Received Date08/21/2019
09/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-