| Model Number |
134046 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Mechanics Altered (2984)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/19/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during distal pancreatectomy, while ligating small vessels around the pancreas, surgeon experienced issues with the loading or firing of the clips.Clips that deployed were malformed.They used another clip applier to complete the case.No patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Post market vigilance (pmv) led an evaluation of one device.The instrument was received partially applied with twelve remaining clips.Visual inspection of the instrument revealed no abnormalities.Functionally, the instrument was fired once in air and proper clip formation was confirmed.The remaining clips were then fired on test media with proper formation.The jaw and handle moved smoothly through the firing cycle and returned to the open position and each remaining clip loaded properly in the jaw.When the cartridge was empty, the interlock engaged and prevented the jaw from approximating.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|