Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation - evaluation reviews of the complaint history, device history record, documentation, manufacturer¿s instructions, and quality control procedures and a visual inspection of the device were conducted during the investigation.The complaint device was not returned.Photos of the device was provided, however, which showed a hair-like fiber present inside the packaging.A document-based investigation evaluation was also performed.There is no evidence to suggest the product was not otherwise made to specifications.A review of the device history record shows no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Furthermore, reviews of the manufacturer¿s instructions and quality control procedures were conducted, and no gaps were discovered.Based on the information provided and no product returned, investigation has concluded that the failure mode was related to the manufacturing and packaging of the device.Measures are being conducted to address this failure mode.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment, no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|