Device is a combination product.Device evaluated by manufacturer: returned product consisted of an agent drug coated balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined, which revealed multiple kinks along the hypotube.The hypotube is separated 35.2cm from the hub.There is blood present on the balloon folds.The balloon is tightly folded.
|
Reportable based on the product analysis completed on june 12, 2019.It was reported that a kink in the shaft and difficulty advancing occurred.A 2.50 mm x 15.00 mm agent balloon was advanced to the mildly calcified and mildly tortuous target lesion.During advancing the agent, it was noticed that the shaft was very soft, very kinked, and almost impossible to advance.The procedure was completed with another of the same device with no issues.No patient complications were reported in relation to this event.However, returned device analysis revealed a separation of the shaft.It was later reported the shaft broke during the procedure.
|