The reason for this revision surgery was reported as an patella tendon tear.The previous surgery and the revision detailed in this investigation occurred 19 days apart.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.Initial or prolonged hospitalization was required.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the event.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.The root cause of this complaint was a revision surgery due to patella tendon tear.Agent has clearly mentioned that "patient fell and had a patella tendon tear", it seems that the event may have possibly occurred due to lack of post-operative care, patient noncompliance with medical instructions, patient activities or trauma.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
|