Model Number 740 |
Device Problems
Defective Alarm (1014); Pressure Problem (3012)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 06/12/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Medtronic has received the suspect device/component from the customer for further evaluation.Upon completing the investigation, the results will be submitted within a supplemental mdr.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the 740 ventilator had a positive end expiratory pressure fault and failed to alarm.There was no patient involvement at the time of the event.
|
|
Manufacturer Narrative
|
(b)(4).Device evaluation summary: the medtronic field service engineer evaluated the ventilator and hoses relevant to the peep system were cleaned.All testing passed per manufacturer specifications.The ventilator was returned to the customer.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Originally it was reported that the 740 ventilator had a positive end expiratory pressure fault and failed to alarm.However, additional information was received confirmed that the ventilator actually did alarm.Both the visual and audible alarms were generated.
|
|
Search Alerts/Recalls
|