| Catalog Number |
K09A |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/10/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent kyphoplasty surgery at t8 due to myeloma.Intra-op, inflatable bone tamp burst.Surgeon started the procedure and all the steps were followed according to the surgical technique, used the drill to remove all the sharp pieces of bone and after that took first balloon and inserted it into the working channel "t15d" and after complete insertion (both the markers were outside the working channel at least 5 mm), started inflating the ibt and after 4 turns the ibt burst inside the vertebral body.Surgeon had to inject the cement like veretebroplasty.There were no patient symptoms or complications as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d10, h3, h6 h6: product analysis results: visual and functional inspection revealed the balloon has been ruptured.The rupture is a slice/cut near the middle of the balloon.This rupture is consistent with the balloon coming in contact with bone spurs when the balloon is inflated in the vertebral body.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|