|
Suspected device evaluated by ok biotech and calculated that the meter operated within specifications we tested the standby current of returned meter, the result was 12.3¿a.The criteria is <55¿a.Pass.Meter setting, audio and all buttons function are ok.We tested the suspected meter with in house control solution and returned strips (strip lot number: d180713-1).The control solution tests for level low were 61/62 mg/dl, for level high were 269/270 mg/dl.The request control solution ranges are: level low 35~85 mg/dl; level high 230~340 mg/dl.All results were within the acceptance range.Pass.We tested the retain strips from our warehouse (same as patient's strip, lot number: d180713-1).The control solution tests for level low were 60/60 mg/dl; for level high were 269/267 mg/dl.The request control solution ranges are: level low 35~85 mg/dl; level high 230~340 mg/dl.All results were within the acceptance range.Pass.
|