• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DJO, LLC CHATTANOOGA INTELECT TRANSPORT COMBO PKG US STD; STIMULATOR, ULTRASOUND AND MUSCLE, FOR USE IN APPLYING THERAPEUTIC DEEP HEAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DJO, LLC CHATTANOOGA INTELECT TRANSPORT COMBO PKG US STD; STIMULATOR, ULTRASOUND AND MUSCLE, FOR USE IN APPLYING THERAPEUTIC DEEP HEAT Back to Search Results
Model Number 2738
Medical Device Problem Codes Electrical /Electronic Property Problem (1198); Sparking (2595)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); Patient Problem/Medical Problem (2688)
Date of Event 06/14/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No device was returned for evaluation.If the device is received, a follow-up report will be submitted upon completion of product evaluation.
 
Event or Problem Description
It was reported that the unit was "not always turning on or shutting off unexpectedly.The unit has a power surge and causing a spark at the electrode, no patient was injured." this reportedly occurred during treatment.No further information was provided.
 
Additional Manufacturer Narrative
E2, e3: unknown.H3, h6: device was returned for evaluation.The product met specifications and did not malfunction per device evaluation.Unit and applicator passed functional testing.No leadwires or electrodes were sent in for evaluation.A new leadwire kit will be sent to the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHATTANOOGA INTELECT TRANSPORT COMBO PKG US STD
Common Device Name
STIMULATOR, ULTRASOUND AND MUSCLE, FOR USE IN APPLYING THERAPEUTIC DEEP HEAT
Manufacturer (Section D)
DJO, LLC
1430 decision street
vista CA 92081 9663
MDR Report Key8775280
Report Number9616086-2019-00049
Device Sequence Number1977411
Product Code IMG
UDI-Device Identifier00888912021883
UDI-Public00888912021883
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K040285
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2738
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/14/2019
Supplement Date Received by Manufacturer06/14/2019
Initial Report FDA Received Date07/10/2019
Supplement Report FDA Received Date07/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-