| Model Number |
2738 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Sparking (2595)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Patient Problem/Medical Problem (2688)
|
| Date of Event |
06/14/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No device was returned for evaluation.If the device is received, a follow-up report will be submitted upon completion of product evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that the unit was "not always turning on or shutting off unexpectedly.The unit has a power surge and causing a spark at the electrode, no patient was injured." this reportedly occurred during treatment.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
E2, e3: unknown.H3, h6: device was returned for evaluation.The product met specifications and did not malfunction per device evaluation.Unit and applicator passed functional testing.No leadwires or electrodes were sent in for evaluation.A new leadwire kit will be sent to the customer.
|
| |
|