|
Pma/510(k) # k083330.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.1 unit of lot c1580448 of echo-25 were returned opened in its original packaging.The device related to this occurrence underwent a laboratory evaluation on 12 june 2019.It was not possible to advance the needle or reinsert the stylet.The needle was removed from the handle during lab evaluation and observed to be crumpled within the handle.Prior to distribution, all echo-25 devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for echo-25 of lot number c1580448 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1580448.The notes section of the instructions for use, which accompanies this device instructs the user to; "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".There is no evidence to suggest that the customer did not follow the instructions for use.The failure of needle kinked/bent was observed in the laboratory.A definitive root cause for the customer complaint could not be determined as circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to hard lesion.Complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
|