According to the reporter, during a cardioverter defibrillator lead implant, the needle broke off when starting to stitch.It was also stated that the device fell into the patient¿s cavity and was removed with an instrument.Another device was opened to complete the case.There was no patient injury.
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of eleven device.A tactile and microscopic inspection of the sealed sutures was performed with no abnormalities.The visual inspection of the sample noted one suture and a needle swage attached.The needle swage was received broken off from the rest of the un-returned needle.Upon microscopic inspection, instrument marks were not observed near the break site.Functionally, a bend moment test was performed on the undamaged needles and met specifications for the product.A review of the device history record indicates this product was released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|