| Model Number |
11484001 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer¿s report of a leak from the silicone tubing of the pump segment was confirmed.Visual inspection of the set noted a small at the top of the silicone segment near the upper fitment.There were no other anomalies observed.Functional and pressure testing confirmed leaking from a small hole at the top of the silicone segment near the upper fitment.The source of the leak is due to a small hole damage in the silicone pump segment near the upper fitment.The root cause of the hole damage was not identified.
|
| |
|
Event or Problem Description
|
|
This product was received unexpectedly and upon investigation was found to have a leak.Although requested, no additional information was received.
|
| |
|
Search Alerts/Recalls
|