C.R. BARD, INC. (COVINGTON) -1018233 DAVOL SILICONE ROUND SINGLE DRAIN; WOUND DRAIN Back to Search Results
Catalog Number 0070320
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Patient Problem/Medical Problem (2688)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown therefore the device history record could not be reviewed.The instructions for use were found adequate and state the following: ¿when placing drain(s) care should be taken to ensure that the perforated portion of the wound drain lies completely within the confines of the wound.Read product insert provided with the closed wound evacuator for detailed instructions, warnings and precautions associated with the use of the evacuator device.To avoid the possibility of drain damage or breakage: additional perforations should not be made in the drains.Avoid suturing through drains.Drains should lie flat and in line with the skin exit areas.Particular care should be taken to avoid any obstacles to the drain exit path.Drains should be checked for free motion during closure to minimize the possibility of breakage.Drain removal should be done gently by hand.They should not be handled with pointed, toothed or sharp instruments which could cause cuts or nicks and lead to subsequent structural failure of the drain.Surgical removal may be necessary if drain is difficult to remove or breaks.Drain placement.Using a single silicone drain: place perforated wound drain within critical fluid collection area of wound.Draw non-perforated section of wound drain through to the outside until drain indicator mark appears at the skin surface.Two sets of indicator marks aid placement of the drain.Trim non-perforated section of drain to desired length.Insert connecting tube of drain to drain port of evacuator.Using two silicone drains: repeat steps 1-3 above.Attach proximal ends of silicone drains to the bard® y-connector (#0070790) and connect y-connector to evacuator drain port.Using a silicone double drain: draw drain using trocar from outside into inside of wound then from inside to outside of wound on other side.Cut wound tube in the middle of perforated section.Remove trocar only by cutting the drain tubing one inch from end of trocar.Trim non-perforated section of drain to desired length.Attach drains to individual evacuators, or to y-connector 0070780 or 0070790.Insert other blue adapter into y-connector and attach drains to each blue adapter.Insert connecting tube into evacuator." (b)(4).
 
Event or Problem Description
It was reported that the wound drain broke half-way through the removal process with part of it still in the patient after approximately 2-months of use.Patient was taken to surgery to have the remainder of the drain removed.Allegedly, the doctor had to "tease away" tissue in-growth from multiple locations with in the drain.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAVOL SILICONE ROUND SINGLE DRAIN
Common Device Name
WOUND DRAIN
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8776770
Report Number1018233-2019-03830
Device Sequence Number13947377
Product Code GBX
UDI-Device Identifier00801741049477
UDI-Public(01)00801741049477
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 07/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0070320
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2019
Initial Report FDA Received Date07/10/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-