Medtronic received information that the tip of two pipelines were not opening.Prior to the event, the marksman was positioned into place and the it was adjusted and riveted and withdrawn to deploy the pipeline and the reported event occurred.The physician repeatedly adjusted the device; however, the tip still would not open.The pipeline was replaced by another pipeline and new marksman and the same event occurred.Considering the patient¿s tortuous vasculature and long duration of the procedure, the operation was completed.It was not known how.The patient was undergoing embolization treatment of an unruptured amorphous aneurysm measuring 20mm x 12mm located in the right internal carotid artery (ica).The distal and proximal landing zone was 4.0mm x 4.2mm.That vasculature was severe in tortuosity.The pipeline was placed in a vessel bend when it failed to open.The marksman was withdrawn and repeatedly adjusted but the issue persisted.Ancillary devices: marksman 150, cook 7f long sheath guide catheter.
|
|
The pipeline flex and marksman catheter were returned.The pipeline flex was found outside the catheter.The distal and proximal dps restraints were found to be intact.The dps sleeves were found intact with no signs of damage.The distal hypotube and ptfe shrink tubing were found to be intact with no signs of elongation.The distal and proximal ends of the pipeline flex braid appeared to be fully opened and moderately frayed.No bend was observed on the pusher.No damages were found with the tip coil, distal marker, re-sheathing marker, resheathing pad or with the proximal bumper.In addition, no damage found the marksman catheter.No other anomalies were observed.Based on the analysis findings, the pipeline flex was not confirmed to have failure to open at the distal end as the distal and proximal ends of the pipeline flex braid were found fully opened and moderately frayed.The damage on the ends of the braid is likely the results of the physician re-sheathing the device more than recommended two times.The customer reported that the patient vessel tortuosity was severe.It is likely that the severe vessel tortuosity may have contributed to the failure to open issue.There was no non-conformance to specifications identified that led to the failure to open issue.Per our instructions for use (ifu): ¿begin to deliver the device using a combination of unsheathing the pipeline flex embolization device and pushing delivery wire simultaneously.After the distal end of the pipeline flex embolization device has successfully expanded, deploy the remainder of pipeline flex embolization device by pushing the delivery wire and/or unsheathing the pipeline flex embolization device.Resheathing and/or manipulation of the micro catheter, by locking down the delivery wire and moving both as a system, may facilitate expansion of the pipeline flex embolization device.The system is designed to allow for 2 full cycles of re-sheathing of the pipeline flex embolization device.Re-sheathing the pipeline flex embolization device more than 2 full cycles may cause damage to the distal or proximal ends of the braid.Do not use in patients in whom the angiography demonstrates the anatomy is not appropriate for endovascular treatment, due to conditions such as severe intracranial vessel tortuosity or stenosis.¿ linked with mdr: 2029214-2019-00715.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|