| Model Number |
MP9251-C |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.The event reportedly occurred in the nicu; therefore it is presumed the patient was a neonate.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a break at the male luer on the maxplus and neutraclear valves stick down when attached to a nicu patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s report that there was a break at the male luer on the maxplus extension set was confirmed by visual inspection.Also observed was stress marks and jagged/uneven edges at the break site.The products were visually inspected for obvious damage such as incomplete bonding engagements, cracks/fractures, kinks, holes, tears and any other anomalies.A residual light brown liquid was observed within the extension set.The male luer fixed collar was broken off from its base.Stress marks and jagged/uneven edges were observed at the break site of the broken collar.Functional testing could not be performed due to the broken collar and the inability to connect it securely to a mating component.Tool marks were present on the mid-body and at the female luer end, and the piston was still recessed, confirming the customer¿s report.No other evidence of damage or anomalies was observed.The root cause was not identified.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a break at the male luer on the maxplus and neutraclear valves stick down, when attached to a nicu patient.Although requested, there is no additional patient or event information available.
|
| |
|