According to the reporter, during laparoscopic cholecystectomy, during firing, the surgeon was able to press the green button but unable to squeeze the handle.It was reported that six reloads broke and fell into the patient's cavity.A grasper was used to remove a piece of tissue with open staples.New handle and reloads were used to complete the case.There was no patient injury.
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but six photos were available for evaluation.Visual inspection noted malformed staples on the cartridge and on tissue.The loading unit anvil was bowed and one of the cartridge of the loading unit was broken.The broken cartridge was in the container.The broken loading unit was found inserted into a trocar and loaded into an instrument.It was reported that the single use loading unit (sulu) was broken, a component disengaged into the cavity, and the instrument did not fire.The reported issues were confirmed.The most likely cause could not be established from the information available.This issue can occur if device was applied over tissue beyond the recommended thickness range.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|