| Catalog Number |
190828 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Reflux within Device (1522)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity.
|
| |
|
Event or Problem Description
|
|
A user facility biomed technician reported to technical support that a 2008k machine experienced a bag of saline filling up.The saline bag backfill occurred during recirculation.The patient was not connected during recirculation.The biomed reported that there were no quick disconnects on the drain line and that the drain line length and height were within specification.The machine was returned to service without a recurrence of the reported issue.An air separator conductivity cable was changed.
|
| |
|
Additional Manufacturer Narrative
|
|
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius regional equipment specialist (res).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
|
| |
|
Search Alerts/Recalls
|