| Model Number |
10607 |
| Medical Device Problem Code |
Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/04/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.Device evaluated by mfr.: synergy ii us mr 2.75 x 38 mm stent delivery system was returned for analysis.A visual examination of the stent found damage to the proximal end of the stent.The first two proximal struts were lifted and pulled distally.The stent showed no signs of movement and was set between the proximal and distal markerbands.The crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were identified.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip found no issues.A visual and tactile examination of the hypotube found multiple kinks along the length of the hypotube shaft.A visual and tactile examination of the outer lumen and the mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 19jun2019.It was reported that crossing difficulties were encountered.The target lesion was located in the highly calcified left anterior descending artery.A 2.75x38mm synergy ii drug-eluting stent was advanced but failed to cross the lesion.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed stent damaged.
|
| |
|
Search Alerts/Recalls
|