• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND 9616671 RECLAIM ASSEM UPPER PULL ROD; RECLAIM INSTRUMENTS : INSERTION DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND 9616671 RECLAIM ASSEM UPPER PULL ROD; RECLAIM INSTRUMENTS : INSERTION DEVICES Back to Search Results
Catalog Number 297500635
Device Problems Break (1069); Scratched Material (3020); Physical Resistance/Sticking (4012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the reclaim assem upper pull rod (d297500635) is jammed and will no longer spin up and down the threads.This tool functioned properly during its last usage in surgery and was discovered to be jammed up after it came through the washer in sterile processing department.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RECLAIM ASSEM UPPER PULL ROD
Type of Device
RECLAIM INSTRUMENTS : INSERTION DEVICES
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI 
MDR Report Key8780843
MDR Text Key150713454
Report Number1818910-2019-98101
Device Sequence Number1
Product Code HWB
UDI-Device Identifier10603295156017
UDI-Public10603295156017
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 06/20/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number297500635
Device Lot NumberSO2030571
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2019
Initial Date Manufacturer Received 06/20/2019
Initial Date FDA Received07/11/2019
Supplement Dates Manufacturer Received08/15/2019
08/15/2019
Supplement Dates FDA Received08/16/2019
08/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-