Erbe electrosurgical generator.Occupation: non-healthcare professional.Investigation evaluation: our evaluation of the product said to be involved determined the cutting wire securing component located near the distal end of the sphincterotome has disconnected from the catheter.The cutting wire is intact and remains securely attached to the sphincterotome at the proximal end.However, due to the catheter and securing component disconnection, the distal end of the cutting wire is no longer connected to the sphincterotome catheter at the distal end.The securing component is intact therefore no part of the device is missing.The cutting wire does shows evidence of cautery application.A product discrepancy or anomaly that could have contributed to this reported occurrence was not observed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Separation of the cutting wire securing component and the catheter can occur if the tip of the sphincterotome is over flexed.The instructions for use caution the user: "do not over flex or bow tip beyond 90 degrees, as this may damage or cause cutting wire to break." other factors that can contribute to separation of the cutting wire securing component and the catheter include manipulating the handle with the catheter in a coiled position or with the precurved stylet inside the cannulating tip.The instructions for use advise the user: "upon removing device from package, uncoil and straighten sphincterotome.Carefully remove precurved stylet from cannulating tip." the instructions for use contain the following comment: "note: do not exercise handle while device is coiled or precurved stylet is in place, as this may cause damage to sphincterotome and render it inoperable." prior to distribution, all fusion pre-loaded with acrobat wire guide sphincterotomes are subjected to a visual inspection and functional test to ensure device integrity.The functional test includes bowing the sphincterotome to ensure the distal end responds to handle manipulation.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
|
|
The following information was reported to the fda on (b)(6) 2019."during an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook fusion pre-loaded with wire guide sphincterotome.The coating stripped from the [pre-loaded] wire guide (see medwatch report 1037905-2019-00286)." on (b)(6) 2019, the device was returned with the cutting wire securing component separated (subject of this report).Our attempts to collect additional information regarding patient outcome were unsuccessful.While the complainant did not specify if the patient experienced any adverse effects or required additional medical procedures due to this event, the information able to be collected does not reasonably suggest the patient was adversely impacted.The additional questions state that no portion of the device detached in the patient.
|