• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD CARIBE LTD. BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN); BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD CARIBE LTD. BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN); BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 365965
Medical Device Problem Code Coagulation in Device or Device Ingredient (1096)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/24/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the sample collected for bilirubin in the bd microtainer® tube with lh (lithium heparin) clotted during use, but was run through the analyzer regardless "with no issues".The following information was provided by the initial reporter: "spoke with customer and they had 2-3 samples that were clotted in the microtainer tubes.They were collected for bilirubin.He feels it is a collection issue by the nurses, that they are not mixing them correctly.Samples were not hemolysed.They did run samples with no issues and have validated the bilirubin test on serum samples.".
 
Additional Manufacturer Narrative
Investigation: investigation summary: bd had not received samples or photos from the customer facility for evaluation; therefore, the investigation was limited.Retention samples were selected from bd inventory for evaluation/testing and upon completion, no issues were observed relating to clotting as all samples met specifications.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Investigation conclusion: based on evaluation of the retain samples, the customer¿s indicated failure mode for clotting with the incident lot was not observed as all samples met the required specifications.Root cause description: based on the investigation, a root cause could not be determined.The retain product was found to be in conformance and meet release specifications.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that the sample collected for bilirubin in the bd microtainer® tube with lh (lithium heparin) clotted during use, but was run through the analyzer regardless "with no issues".The following information was provided by the initial reporter: "spoke with customer and they had 2-3 samples that were clotted in the microtainer tubes.They were collected for bilirubin.He feels it is a collection issue by the nurses, that they are not mixing them correctly.Samples were not hemolysed.They did run samples with no issues and have validated the bilirubin test on serum samples.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MICROTAINER® TUBES WITH LH (LITHIUM HEPARIN)
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos
MDR Report Key8781572
Report Number2618282-2019-00223
Device Sequence Number13962801
Product Code JKA
UDI-Device Identifier50382903659659
UDI-Public50382903659659
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K991702
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2020
Device Catalogue Number365965
Device Lot Number830556N
Initial Date Received by Manufacturer 06/26/2019
Supplement Date Received by Manufacturer06/26/2019
Initial Report FDA Received Date07/11/2019
Supplement Report FDA Received Date08/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-