• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL2¿, 63 CM LENGTH, 8.5 F; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL2¿, 63 CM LENGTH, 8.5 F; INTRODUCER, CATHETER Back to Search Results
Model Number 407455
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cardiac Tamponade (2226); Cardiac Perforation (2513)
Date of Event 06/24/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event or Problem Description
Manufacturing related ref: 3008452825-2019-00329, 3005334138-2019-00371, 3008452825-2019-00328.During the pulmonary vein isolation (pvi) ablation procedure, the patient experienced a pericardial effusion.An ice catheter was used to view the ventricles and pericardium throughout the procedure and the user noted a small effusion developing during the left vein wide atrial circumferential ablation.The effusion was in the left atrium and the patient experienced reduced blood pressure and cardiac tamponade.The procedure was stopped and protamine was administered and a pericardiocentesis was performed with the placement of a pericardial drain.There were no performance issues with an abbott device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL2¿, 63 CM LENGTH, 8.5 F
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
MDR Report Key8781895
Report Number3005334138-2019-00372
Device Sequence Number2540307
Product Code DYB
UDI-Device Identifier05414734205795
UDI-Public05414734205795
Combination Product (Y/N)N
Initial Reporter StateLA
Initial Reporter CountryUS
PMA/510(K) Number
K052644
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/11/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model Number407455
Device Lot Number6955077
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2019
Initial Report FDA Received Date07/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ADVISOR¿ FL, SE¿; AGILIS¿ NXT INTRODUCER; TACTICATH¿ SE¿
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient Weight79
-
-