MEDTRONIC EUROPE SARL PROTECTA DR; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER Back to Search Results
Model Number D334DRG
Medical Device Problem Code Electromagnetic Interference (1194)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/20/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that there was an audible alert from the device due to the patient being too close to a magnet.Follow up concluded no intervention was performed and the patient is now aware to be cognizant when around magnets as a tone will be delivered.The device remains in use.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory indicated oversensing due to electromagnetic interference/noise.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTECTA DR
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
MDR Report Key8782240
Report Number9614453-2019-02321
Device Sequence Number19327286
Product Code LWS
UDI-Device Identifier00613994678157
UDI-Public00613994678157
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P980016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date03/28/2012
Device Model NumberD334DRG
Device Catalogue NumberD334DRG
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/20/2019
Supplement Date Received by Manufacturer07/14/2019
Initial Report FDA Received Date07/11/2019
Supplement Report FDA Received Date07/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
407652 LEAD, 693558 LEAD
Patient Age76 YR
Patient Weight86
-
-