(b)(4).A manufacturing record evaluation was performed for the finished device lot number mgq067, and no non-conformances related to the reported complaint condition were identified.Device evaluation summary: a suture piece, an empty opened foil and an empty opened folder of product code 1744, lot #: mgq067, was returned for evaluation.This product code is double armed.During the visual inspection of the sample, the suture ends were noted cut appears to be by use of a surgical instrument and present body fluids along of the strand.The needles were not returned for evaluation.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of the sample, it could not be determined what may have caused the reported incident.An empty opened box and six representative samples of product code 1744, lot #: mgq067, were returned for evaluation.During the visual inspection of six samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, or damaged were observed.A functional test was performed and the pull force were above the minimum requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the conditions of the samples received, no performance pull off suture needle was found, and the tested samples met the finished goods requirements.Representative samples returned to support investigation of 3 device issues captured in reports 2210968-2019-84119, 2210968-2019-84122 and 2210968-2019-84126.Analysis results have been included in reports for each.
|