| Model Number |
425-2008X |
| Medical Device Problem Codes |
Break (1069); Device-Device Incompatibility (2919); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
For the reported information, the device was returned to bd and evaluated.The result of the investigation is confirmed for the break issue reported.There was a break noted in the shaft 210mm from the strain relief.There was a also a kink noted 15mm from the break and a number of bends along the length of the shaft.The method of packaging may have contributed to these bends and the kink.The definitive root cause for the reported break issue could not be determined based upon available information.The device is labeled for single use.
|
| |
|
Event or Problem Description
|
|
This report summarizes one (1) malfunction.A review of the reported information indicated that model 425-2008x pta dilatation catheter allegedly experienced a break, material deformation, and device-device incompatibility.This information was received from one source.The malfunction involved a patient with no known impact to the patient.Patient age, weight, and gender were not provided.
|
| |
|
Search Alerts/Recalls
|