• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL, L.P. EMPOWR 3D KNEE; EMPOWR 3D KNEETM INS, 6R 16MM, VE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL, L.P. EMPOWR 3D KNEE; EMPOWR 3D KNEETM INS, 6R 16MM, VE Back to Search Results
Model Number 342-16-706
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 06/12/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as an infection.The previous surgery and the revision detailed in this investigation occurred 49 days apart.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.Initial or prolonged hospitalization was required.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhrs) show that the reported component used in the previous surgery met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to an infection.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The device was verified to have gone through an acceptable sterilization process and was within it's expiration date at the time of use during the previous surgery.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's infection.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - patient acquired a post-op infection.Surgeon performed a wash out and went up one size on the poly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR 3D KNEE
Common Device Name
EMPOWR 3D KNEETM INS, 6R 16MM, VE
Manufacturer (Section D)
ENCORE MEDICAL, L.P.
9800 metric blvd.
austin texas 78758
MDR Report Key8783129
Report Number1644408-2019-00631
Device Sequence Number19249617
Product Code JWH
UDI-Device Identifier00888912167635
UDI-Public(01)00888912167635
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K143242
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/01/2022
Device Model Number342-16-706
Device Catalogue Number342-16-706
Device Lot Number160T1013
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2019
Initial Report FDA Received Date07/11/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age77 YR
-
-