The device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.The device for the malfunction has not been returned for evaluation; therefore, the investigation is inconclusive for material rupture as no objective evidence has been provided to confirm any alleged deficiency with the device.Based upon the available information, the definitive root cause is unknown.The device is labeled for single use.The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
This report summarizes one malfunction.A review of the reported information indicated that model 425-3004x rx pta catheter allegedly experienced a material rupture.This information was received from a single source.The malfunction involved a patient with no consequence.The patient¿s age, weight, and sex were not provided.
|