BARD PERIPHERAL VASCULAR, INC. FLUENCY PLUS ENDOVASCULAR STENT GRAFT Back to Search Results
Model Number FEM08100
Medical Device Problem Codes Positioning Failure (1158); Fracture (1260); Material Perforation (2205)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
For the reported information, the device was returned to bd for evaluation.The company is still investigating the issue at this time.The device is labeled for single use.
 
Event or Problem Description
This report summarizes one (1) malfunction.A review of the reported information indicated that model fem08100 endovascular stent graft allegedly experienced failure to deploy, a fracture, and material perforation.This information was received from one source.The malfunction involved a patient with no known impact to the patient.The male patient's age and weight were not provided.
 
Additional Manufacturer Narrative
H10: for the reported information, the lot number was provided and a lot history review was performed.As a result of the investigation performed the complaint is confirmed for a positioning failure, fracture, and material perforation.A definite root cause for the reported event could not be determined.The device is labeled for single use.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
This report summarizes one (1) malfunction.A review of the reported information indicated that model fem08100 endovascular stent graft allegedly experienced failure to deploy, a fracture, and material perforation.This information was received from one source.The malfunction involved a patient with no known impact to the patient.The male patient's age and weight were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLUENCY PLUS ENDOVASCULAR STENT GRAFT
Common Device Name
ENDOVASCULAR STENT GRAFT
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8783335
Report Number2020394-2019-01480
Device Sequence Number19467900
Product Code PFV
UDI-Device Identifier04049519008516
UDI-Public(01)04049519008516
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberFEM08100
Device Catalogue NumberFEM08100
Device Lot NumberANCY4237
Initial Date Received by Manufacturer 06/30/2019
Supplement Date Received by Manufacturer09/30/2019
Initial Report FDA Received Date07/11/2019
Supplement Report FDA Received Date10/09/2019
Patient Sequence Number1
-
-