MENTOR TEXAS MENTOR SILTEX CONTOUR PROFILE SPECTRUM; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE Back to Search Results
Catalog Number 3542512
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 06/11/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a (b)(6) year-old caucasian female patient underwent a breast augmentation primary with a mentor siltex contour profile spectrum 350cc and experienced right side deflation.As a result, the patient had undergone removal and replacement with a mentor siltex contour profile spectrum 350cc on (b)(6) 2019.
 
Additional Manufacturer Narrative
On (b)(6) 2019, it was reported to mentor that the replacement device was mentor siltex contour profile high 350cc.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On 08/29/2019, the mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On 9/18/2019, during evaluation of the sample it was observed to be intact.Leak testing was performed and it revealed a tear in the posterior view, measurement approximately 0.2 cm.Microscopic examination of the edges of the rupture revealed parallel striations.No other anomalies were noted.The striations are consistent with markings made by a sharp instrument perforating silicone material.As stated in the product insert data sheet, do not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps or scissors to contact the device during the implantation or other surgical procedures as this can result in rupture.The manufacturing records were reviewed, and the manufacturing criteria were met prior to the release of this lot.It should be noted that as part of our quality process, each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Based on the information reported and/or the product investigation, there is no evidence that the issue is related with the manufacturing process.Complaint information will be included in complaint trending that is reviewed and monitor by quality assurance on a regular basis to determine if further action is necessary.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SILTEX CONTOUR PROFILE SPECTRUM
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key8783453
Report Number1645337-2019-15228
Device Sequence Number2280872
Product Code FWM
UDI-Device Identifier00081317002147
UDI-Public00081317002147
Combination Product (Y/N)N
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2009
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/23/2013
Device Catalogue Number3542512
Device Lot Number5903914
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2019
Initial Date Received by Manufacturer 06/18/2019
Supplement Date Received by Manufacturer08/29/2019
09/18/2019
11/01/2019
Initial Report FDA Received Date07/11/2019
Supplement Report FDA Received Date08/29/2019
09/18/2019
11/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
-
-