| Catalog Number |
9391326 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
06/06/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product was not returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent spinal fusion surgery at t11/12 and l2/3 due to vertebral body collapse and pseudoarthrosis occurred to l1.Post-op, the patient fell from the bed at the 3rd day after the operation and also tumbled several times.The patient did not listen to the healthcare professional.The sinking and movement of t2, and the load to the screw occurred due to tumbling.Patient suffered pain in lumbar and back and prolonged hospitalization as a result of this event.The implanted cage was deviated from its position.Patient underwent revision surgery as a result of this event.In which, doctor adjusted the position of the cage, and the operation was completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Radiological image review results: post-op x-ray ct for anterior/posterior stabilization of l1 fracture.The anterior interbody graft is deviated and positioned into the interior vertebral body with little baseplate extent with the superior body no anterior fixation plate is present, the base plates appear undersized compared to the vertebral bodies and expanded height of the cage is taller than the corpectomy space.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|