| Model Number |
E SERIES |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
06/21/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The reported malfunction was observed and attributed to the lithium coin battery on the system board.The device was recertified and returned to the customer.Zoll medical corporation recommends replacement of the lithium battery on the system board every 5 years.This device is approximately 9 years old from date of manufacture.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device would not power up.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
|
| |
|