| Medical Device Problem Code |
Positioning Problem (3009)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
|
| |
|
Event or Problem Description
|
|
The customer reported a suspected deviation on dbs position on right brain hemisphere.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation was completed by conducting a thorough evaluation of the product and the reported information.The arc system has been investigated and no mechanical error has been found and the accuracy of the system is within specification.The user has also used the system after mechanical investigation with good results.It is judged that the information and warnings in the instructions for use are sufficient for proper handling of the leksell vantage stereotactic system.
|
| |
|