• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDELA LLC PUMP IN STYLE ADVANCED - ON-THE-GO TOTE; PUMP, BREAST, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDELA LLC PUMP IN STYLE ADVANCED - ON-THE-GO TOTE; PUMP, BREAST, POWERED Back to Search Results
Model Number 57063
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 06/11/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Customer service conducted troubleshooting with the customer, including inspecting the faceplate and back plate for damage; inspecting and cleaning the diaphragm; disassembling, inspecting, cleaning and reassembling the kit and tubing set.The troubleshooting did not resolve the issue.The customer was sent a replacement pump and return of her original pump was requested for testing/evaluation.In follow up with a complaint handler on 06/25/2019, the customer indicated that she developed mastitis on (b)(6) 2019 and was prescribed an antibiotic, which had resolved the mastitis.She additionally indicated that the replacement pump was working without issue.Based on the results of (b)(4), it cannot be definitively concluded that the pump caused or contributed to the customer's mastitis.The estimated incidence of mastitis in lactating women, whether using a breast pump or not, according to published clinical literature can be as high as 33%.In fact, clinical guidelines suggest the use of a breast pump to facilitate withdrawal of breast milk during bouts of mastitis.The complaint rate of mastitis across all reported failures, across all medela breast pumps, is 0.008% for the period of january 2013 to august 2017.Mastitis is usually a benign, self-limiting infection with few consequences for the suckling infant.The risk of mastitis is higher among women who have breastfed previously, especially those with a history or mastitis." riordan & wambach, 4th ed.P.294: breastfeeding and human lactation.Mastitis requires prompt medical attention for the mother for pain relief and prescription antibiotics to avoid progression to overwhelming sepsis.
 
Event or Problem Description
On (b)(6) 2019, the customer alleged to medela llc that she was experiencing low suction with her pump in style breast pump.She additionally alleged that she had mastitis due to the low suction and she had seen her healthcare provider, who prescribed her antibiotics.
 
Additional Manufacturer Narrative
The device was evaluated on 07/15/2019 with the customer's parts as received and it failed vacuum testing.The vacuum test was then conducted using a lab kit with the customer's pump and it passed.Refer to attached product evaluation.When a pump fails with the customer kit, but not the lab kit, any issue is typically attributable to the kit, not the pump.In this case, it was noted in the evaluation that the customer's connectors, tubing and valves had residue on them and the customer's membranes were not laying flat.Any foreign contaminants that hinder the device's ability to form an adequate vacuum seal could potentially result in lower suction.It is very difficult to determine exactly how much contamination under normal use is needed for pump vacuum to suffer, but it is a known failure mode causing low suction.The pump in style instructions for use details cleaning procedure for the parts and accessories.It is also a common troubleshooting item to check.The customer's report of low suction was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP IN STYLE ADVANCED - ON-THE-GO TOTE
Common Device Name
PUMP, BREAST, POWERED
Manufacturer (Section D)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
MDR Report Key8784421
Report Number1419937-2019-00093
Device Sequence Number2398001
Product Code HGX
UDI-Device Identifier00020451570631
UDI-Public00020451570631
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K031614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number57063
Device Catalogue Number57063
Device Lot Number448465
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2019
Initial Date Received by Manufacturer 06/19/2019
Supplement Date Received by Manufacturer07/15/2019
Initial Report FDA Received Date07/12/2019
Supplement Report FDA Received Date07/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-