• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS TOTAL ABSCESSION BILIARY DRAINAGE CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS TOTAL ABSCESSION BILIARY DRAINAGE CATHETER Back to Search Results
Catalog Number H787140009045
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 05/31/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported defective device has yet to be returned to the manufacturer for a device evaluation.The firm is attempting to obtain the device.An investigation into the root cause for product problem is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.(b)(4).
 
Event or Problem Description
The same patient has had multiple internal/external biliary drainage catheters disintegrate.They were placed on (b)(6) 2018, (b)(6) 2018, (b)(6) 2018, (b)(6) 2019, and (b)(6) 2019.All were angiodynamics 14fr 40cm biliary drain total abscession catheters.I asked if they were using anything corrosive to clean.The surgeon spoke w/ the pt.& her family and they state that they are flushing out the tube only w/ saline.While the pt.Went more than 4 weeks in between changes this time around, the surgeon said that at times he has seen deterioration in the tubing in as little as 3 weeks in between changes, in the past.Md attempted to remove as many drain pieces as he could.Pigtail was left in the patient.It was reported the defective disposable device is available for return to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
Returned for evaluation were two partial drainage catheters.The drainage catheter was returned not in its entirety.Examination by the manufacturing engineer for this product determined there was no evidence of material stretching or embrittlement at the site of the catheter fracture.This failure mode is consistent with excessive force being applied to the suture, most likely the result of difficulty in extraction of the device.The customer's reported complaint description of "the drainage catheter broke off" is confirmed.The catheter tubing was observed to be within dimension specification and there was no evidence of material stretching, nor were there any obvious visual indications of damage during manufacture that could have led to compromised integrity.The cause of the difficulty cannot be conclusively ascertained, but the failure mode is consistent with material breakdown due to exposure to an incompatible substance, possibly alcohol.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.The instructions for use (ic 188) which is supplied to the end user with this catalog number states, "warnings: do not use this catheter with alcohol.Placing a 10 french or larger catheter as the primary drain before formation of a tract may be difficult in some patients.In these patients, the initial biliary drainage should be started with a smaller (8 french) catheter until a suitable tract allows placement of a larger catheter.Where long-term use is indicated, it is recommended that indwelling time not exceed 90 days.This catheter should be evaluated by the physician on or before 90 days post placement." a review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOTAL ABSCESSION BILIARY DRAINAGE CATHETER
Common Device Name
TOTAL ABSCESSION BILIARY DRAINAGE CATHETER
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
MDR Report Key8784798
Report Number1319211-2019-00066
Device Sequence Number16561176
Product Code DQO
UDI-Device IdentifierH787140009045
UDI-PublicH787140009045
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
K002101
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/23/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2021
Device Catalogue NumberH787140009045
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/07/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2019
Supplement Date Received by Manufacturer06/25/2019
Initial Report FDA Received Date07/12/2019
Supplement Report FDA Received Date10/23/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
-
-