• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC NAVLOCK BLUE; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC NAVLOCK BLUE; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9736061
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 05/30/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation of the returned tracker determined that it had areas of galling inside the handle causing the reported fit issues.Otherwise, with markers attached and fully seated, the tracker returned a good geometry error with normal tracking.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information from a healthcare professional (hcp) via a manufacturer representative (rep) regarding a guidance device being used in a spinal procedure.It was reported that both the blue and black navlocks were sticking to the instruments not allowing them to spin freely.There was no patient involvement.There were no additional complications reported or anticipated as a result of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVLOCK BLUE
Common Device Name
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key8785078
Report Number1723170-2019-04058
Device Sequence Number2038228
Product Code HAW
UDI-Device Identifier00763000123185
UDI-Public00763000123185
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K050438
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/12/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9736061
Device Catalogue Number9736061
Device Lot Number181024
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/13/2019
Initial Report FDA Received Date07/12/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-