DEPUY INTERNATIONAL LTD. 8010379 AML 12.0 MMA 6.3 5/8 STD 12/14; AML IMPLANT : HIP FEMORAL STEM Back to Search Results
Catalog Number 155703000
Medical Device Problem Codes Fracture (1260); Loss of Osseointegration (2408)
Health Effect - Clinical Codes Inadequate Osseointegration (2646); No Code Available (3191)
Date of Event 06/21/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that this 5/8 coated aml stem fractured proximally where the metaphyseal portion tapered down to the straight cylinder diaphyseal part.The stem was removed.The cup was retained and a new liner inserted.It was noted that there was loosening of the stem at the bone to implant interface.Doi unknown; dor: (b)(6) 2019; left hip.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other related reported incidents against the provided product code/lot number combination since release for distribution.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Device history lot: t3yc51.Device history batch: null.Device history review: the 5 addenda were for reboxing of individual part.For the dhr no other deviations or anomalies were observed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AML 12.0 MMA 6.3 5/8 STD 12/14
Common Device Name
AML IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD. 8010379
st. anthony's rd
leeds LS11 8 DT
UK  LS11 8 DT
MDR Report Key8785130
Report Number1818910-2019-98172
Device Sequence Number1690664
Product Code LPH
UDI-Device Identifier10603295059240
UDI-Public10603295059240
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K953694
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number155703000
Device Lot NumberT3YC51
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/21/2019
Supplement Date Received by Manufacturer08/01/2019
Initial Report FDA Received Date07/12/2019
Supplement Report FDA Received Date08/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN HIP ACETABULAR LINERS; UNKNOWN HIP FEMORAL HEAD; UNKNOWN HIP LOCKING RING; UNKNOWN HIP ACETABULAR LINERS; UNKNOWN HIP FEMORAL HEAD; UNKNOWN HIP LOCKING RING
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight85
-
-