It was reported that after approximately 45 minutes after initiation on va ecmo, a low flow alarm was triggered.When healthcare providers we looked at the screen, it said ¿no flow¿.Healthcare providers emergently did a pump change and restored the revolutions per minute(rpm) and flow measured in liters per minute(lpm) back to what they were before the no flow alert occurred.Healthcare providers weren¿t aware of anything that happened that would cause that sudden failure in flow.No additional information was provided.
|
Manufacturer's investigation conclusion: the report of "no flow" and a speed reading of 0rpm was confirmed through the analysis of a data log file retrieved from the returned centrimag 2nd gen primary console (sn (b)(6), evaluated under (b)(4) associated with this event.Per the log file, the console was powered on at approximately 2:57pm on (b)(6) 2019 and the system was initially set to support a pump at a set speed of ~1900rpm and a flow rate of ~2.8lpm and the system operated without any issues.At approximately 3:11pm speed was adjusted to ~4200rpm and then increased to ~4400rpm.After this speed adjustment flow was captured at ~4.6lpm.At approximately 3:20pm the log file captured a momentary flow below minimum:f3 alert during which flow was captured at 0.007lpm.This resolved within 10 seconds.At ~3:31pm speed was increased to ~5000rpm and flow increased to ~5.5lpm.The system operated on ac power without any issues until approximately 4:03pm, when a shutdown event was captured.Following this event speed dropped to 0rpm and flow was measured at ~-1.0lpm and the console alarmed with motor disconnected:m2 and flow signal interrupted:f2 alerts.At ~4:04pm the pump was captured as disconnected and the console was powered down at ~4:10pm.The returned centrimag motor (sn (b)(6) was evaluated and tested by the service depot.The reported complaint could not be verified nor duplicated during their evaluation.The returned motor was operated with its associated 2nd gen primary console (sn (b)(6), evaluated under mfr#2916596-2019-03185) and flow probe (sn (b)(6), evaluated under mfr#2916596-2019-03335) for an extended period of time.The motor was also tested with a test console and flow probe.No alarms nor any other issues were observed.The returned motor performed as intended.Visual inspection of the motor's cable did not reveal any issues.The motor was functionally tested per the centrimag motor service process and the unit passed all tests.The motor was found to function as intended.As a result, the root cause of the reported event could not be conclusively determined nor correlated to a motor related issue.Reports of similar events will continue to be tracked and monitored.No further information was provided.The manufacturer is closing the file on this event.
|