| Catalog Number |
04P52-01 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation into this issue determined that there is a high possibility that control and calibrator materials had been mixed during the fill process and that hba1c calibrator lot 54582uq02 may contain a quantity of high control material instead of calibrator level 2 (l2).A product recall letter was issued to all architect customers who have received the architect hba1c calibrator lot 54582uq02.This lot may generate an active calibration curve, a shift to qc and patient results may occur.The letter instructs.
|
| |
|
Event or Problem Description
|
|
The customer reported a calibration failure while using the architect hba1c calibrator lot 54582uq02.There was no report of incorrect patient results or impact to patient management.
|
| |
|