| Model Number |
FEM12040 |
| Medical Device Problem Codes |
Positioning Failure (1158); Misfire (2532)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device for this malfunction has been returned to the manufacturer for evaluation.The investigation of the reported malfunction is currently underway.The device is labeled for single use.
|
| |
|
Event or Problem Description
|
|
This report summarizes one malfunction.A review of the reported information indicated that model fem12040 endovascular stent graft allegedly experienced a failure to deploy and a misfire.This information was received from a single source.The malfunction involved a patient with no reported patient consequence.The patient was reported as a male weighing (b)(6) lbs.The patient¿s age was not reported.
|
| |
|
Event or Problem Description
|
|
This report summarizes one malfunction.A review of the reported information indicated that model fem12040 endovascular stent graft allegedly experienced a failure to deploy and a misfire.This information was received from a single source.The malfunction involved a patient with no reported patient consequence.The patient was reported as a male weighing 190 lbs.The patient¿s age was not reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the lot number for the malfunction was provided and a lot history review was performed.The device has been returned for evaluation; the evaluation confirmed the device partially deployed.A definite root cause for the reported event could not be determined.The device is labeled for single use.H11: section a through f: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|