ZOLL MEDICAL CORPORATION X SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number X SERIES
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The malfunction was duplicated and the monitor board was replaced to resolve the malfunction.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
 
Event or Problem Description
Complainant alleged that during biomed testing, the device would not power up.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key8787404
Report Number1220908-2019-01933
Device Sequence Number1999188
Product Code MKJ
UDI-Device Identifier00847946012717
UDI-Public00847946012717
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K112432/P160
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/05/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2019
Initial Report FDA Received Date07/13/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-